Human Reproduction, Vol. 14, No. 5, 1175-1177,
May 1999
© 1999 European Society of Human Reproduction and Embryology
Precision and method bias of two assays for oestradiol: consequences for decisions in assisted reproduction
1 Department of Obstetrics and Gynaecology, The University of Western Ontario and 2 Gamma-Dynacare Medical Laboratories, London, Ontario, Canada
Oestradiol monitoring in assisted reproduction is empirical, with no proof of benefit. Precise oestradiol estimation will be an essential pre-requisite for high quality evaluation of possible differences between combined and ultrasound-only monitoring. Objectives of the present trial were independent method comparison and bias estimation of chemiluminescent immunoassay (CLIA) versus radioimmunoassay for oestradiol. In a prospective comparison, 505 consecutive samples were split and assayed concurrently. Precision (reproducibility), relative bias and logistics were analysed and compared to manufacturers' findings. Correlation between CLIA and radioimmunoassay was excellent. Positive bias with CLIA necessitated altering decision points for therapy. Precision (reproducibility) was superior with CLIA, making it an appropriate candidate method for future randomized trials of the effectiveness of combined oestradiol/ultrasound monitoring for assisted reproduction.
Key words: bias/immunoassay/IVF/oestradiol/precision
3 To whom correspondence should be addressed at:339 Windermere Road, London, Ontario, Canada N6A 5A5
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
A. S. Dighe and P. M. Sluss Improved Detection of Serum Estradiol after Sample Extraction Procedure Clin. Chem., April 1, 2004; 50(4): 764 - 766. [Full Text] [PDF] |
||||
![]() |
J. Taieb, C. Benattar, R. Diop, A. S. Birr, and A. Lindenbaum Use of the Architect-i2000 Estradiol Immunoassay during in Vitro Fertilization Clin. Chem., January 1, 2003; 49(1): 183 - 186. [Full Text] [PDF] |
||||
![]() |
E. H. Yu Ng, W. S. B. Yeung, E. Yee Lan Lau, W. W. K. So, and P. C. Ho High serum oestradiol concentrations in fresh IVF cycles do not impair implantation and pregnancy rates in subsequent frozen-thawed embryo transfer cycles Hum. Reprod., February 1, 2000; 15(2): 250 - 255. [Abstract] [Full Text] [PDF] |
||||