Hum. Reprod. Advance Access published online on May 19, 2005
Human Reproduction, doi:10.1093/humrep/dei061
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
1 Unidad de Reproducción, Hospital Virgen de las Nieves, E-18014, Granada, Programa de Control de Calidad Externo de Análisis de Semen, Asociación para el Estudio de la Biología de la Reproducción (ASEBIR)
* To whom correspondence should be addressed. BACKGROUND: The aim of this study was to calculate the analytical goal for seminal parameters based on the state of the art, and then to compare these specifications with those previously obtained by our group based on biological variation. METHODS: All data used for analysis were derived from the Spanish programme of external quality control on semen analysis. Over 90 laboratories participated from 1999 to 2003. Using graphs of the state of the art, we also determined the numbers of laboratories that achieved quality specifications. RESULTS: The total allowable error calculated using state of the art graphs is similar to that calculated using biological variation for concentration and total motility. However, it is much higher for morphology and rapidly progressive motility. Over 80% of the laboratories achieved the minimum quality specification based on biological variation for concentration, total and progressive motility. However, only
Received December 17, 2004
Revised April 4, 2005
Accepted April 7, 2005
Article
Quality specifications for seminal parameters based on the state of the art
2 Programa de Supervisión Externa de la Calidad, Asociación Española de Farmacéuticos Analistas, AEFA, Madrid
3 Programa de Control de Calidad Externo de Análisis de Semen, Asociación para el Estudio de la Biología de la Reproducción (ASEBIR), Unidad de Reproducción, Clínica Avicena, Jaén
4 Programa de Control de Calidad Externo de Análisis de Semen, Asociación para el Estudio de la Biología de la Reproducción (ASEBIR), CEIFER, Granada
5 Programa de Control de Calidad Externo de Análisis de Semen, Asociación para el Estudio de la Biología de la Reproducción (ASEBIR), Departamento de Ciencias de la Salud, Universidad de Jaén, Jaén, Spain
6 Unidad de Reproducción, Hospital Virgen de las Nieves, E-18014, Granada
J.A. Castilla, E-mail: josea.castilla.sspa{at}juntadeandalucia.es
![]()
Abstract
30% of the laboratories achieved the minimum quality specification based on biological variation for morphology and rapidly progressive motility. CONCLUSIONS: The study enabled us to identify the state of the art of analytical performance for seminal parameters, and revealed the difficulty inherent in meeting the quality specifications based on biological variation.![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
J. A. Castilla, J. Hernandez, Y. Cabello, A. Lafuente, N. Pajuelo, J. Marqueta, B. Coroleu, and (Assisted Reproductive Technology Register of the Defining poor and optimum performance in an IVF programme Hum. Reprod., January 1, 2008; 23(1): 85 - 90. [Abstract] [Full Text] [PDF] |
||||
![]() |
J.A. Castilla, C. Alvarez, J. Aguilar, C. Gonzalez-Varea, M.C. Gonzalvo, and L. Martinez Influence of analytical and biological variation on the clinical interpretation of seminal parameters Hum. Reprod., April 1, 2006; 21(4): 847 - 851. [Abstract] [Full Text] [PDF] |
||||